Abstract:The concentration of 40K in seawater was measured directly by HPGe γ spectrometer. Based on the LabSOCS software, the maximum detection efficiency corresponding to the optimal geometric shape and parameters was calculated, as a result, the analysis of the concentration of 40K in seawater can be finished in 3 hours with relative standard deviation less than 10% for sample net count. Compared with other methods, γ-spectrum analysis has few influencing factors, and it does not need any pretreatment process which is very convenient. Therefore, the γ-spectrum analysis is worth promoting as an alternative method in environmental radiological baseline study.
CANBERRA.User's manual of model S573 ISOCS calibration software[M].USA,2002.
[4]
SIMA O,ARNOLD D.Self-attenuation and coincidence-summing corrections calculated byMonte Carlo simulations for gamma-spectrometric measurements with well-type germanium detectors[J].Applied Radiation and Isotopes,1996,47(9/10):889-893.
[5]
RUSS W,VENKATARAMAN R,BRONSONF.Validation testing of the genie 2000 Cascade Summing Correction[J].Journal of Radioanalytical and Nuclear Chemistry,2005,264(1):193-197.
BRONSON F L.Validation of the accuracy of the LabSOCS software for mathematical efficiency calibration of Ge detectors for typical laboratory samples[J].Journal of Radioanalytical and Nuclear Chemistry,2003,255(1):137-141.
AHMED A S,CAPELLO K,CHIANG A,et al.Optimization of geometric parameters for Marinelli beaker to maximize the detection efficiency of an HPGe detector[J].Nuclear Instruments and Methods in Physics Research Section A:Accelerators,Spectrometers,Detectors and Associated Equipment,2009,610(3):718-723.